Directives
As a first step, you need to verify whether your product(s) fall(s) within the definition of a medical device in accordance with Article 1 paragraph 2a) of Directive 93/42/EEC. Second, you need to exclude that your product(s) falls within the definition of an active implantable medical device (Directive 90/385/EEC) or of an in vitro diagnostic medical device (Directive 98/79/EC). Finally, you must verify that no other exclusion clause contained in the first Article of Directive 93/42/EEC is applicable. If all these conditions are fulfilled, Directive 93/42/EEC applies.
Verify requirements
The New Approach Directives for the CE marking have been designed by the European Union in such a manner that they cover, within their respective scope, all requirements for products from each of the aforementioned sectors.
However, it is possible that more than one New Approach Directive applies to the same product. Furthermore, other legislation (such as horizontal legislation on chemicals or environment) can apply.
Directive 93/42/EEC on Medical Devices (MDD) specifies general requirements that the product has to meet in order for the manufacturer to affix the CE marking. These are the so-called Essential Requirements that are listed in Annex I to the MDD. Compliance with the essential requirements must be demonstrated by a clinical evaluation in accordance with Annex X to Directive 93/42/EEC.
Need for notified body?
Before proceeding with the assessment procedure, it is important to establish whether you, the manufacturer, can assess your product by yourself or whether you have to involve a Notified Body.
In the Directive on Medical Devices, the medical device is classified in accordance with the provisions of Annex IX of the MDD as class I (low risk), class IIa or IIb (medium risk) or class III (high risk). The involvement of a Notified Body is not necessary for medical devices of class I unless they have a measuring function or are placed on the market in a sterile condition.
A Notified Body verifies in most of the cases only the quality management of the manufacturer. For all medical devices belonging to class III, and for medical devices belonging to class IIa and IIb on a representative basis, the design of the medical device and its compliance with the Essential Requirements must be examined by a Notified Body. The Notified Body issues a certificate that indicates, by referring to one of the Annexes II to VI of the MDD, what has been verified.
To find the Notified Bodies appointed by the Member States to carry out conformity assessment, please consult NANDO – the New Approach Notified and Designated Organisations database. You can search for Notified Bodies by Directive, country or by the Notified Body's number (indicated next to the CE marking) via the NANDO homepage.
Check conformity
Subject to the class of the medical device, the manufacturer has different options as to how to assess the conformity of the medical device. The stringency of the conformity assessment procedures depends on the class of the medical device. For each class, the manufacturer has the choice between two or more conformity assessment procedures. Each of the conformity assessment procedures consists of the application of one or more Annexes out of Annexes II to VII of the MDD. In both cases, the clinical evaluation must be part of the documents which the manufacturer submits to the Notified Body for assessment and the Notified Body issues (a) certificate(s) that indicate(s), by referring to one of the Annexes, what has been verified.
Whether the involvement of a Notified Body has been necessary or not, the manufacturer must draw up the Declaration of Conformity (DoC) to declare his sole responsibility for the conformity to the relevant Directive. The Declaration of Conformity must include the manufacturer’s details such as name and address, essential characteristics of the product, if applicable the identification number of the Notified Body as well as a legally binding signature on behalf of the organisation.
Technical documentation
Prior to submission of an application to the Notified Body or, at the latest, when placing the first device of the respective type on the market, the manufacturer must establish the technical documentation. The technical documents must enable the assessment of the conformity of the appliance with the requirements of the Directive.
The manufacturer or the authorised representative established in the European Community is requested to keep copies of the technical documentation for a period of at least 5 years, in the case of implantable devices at least 15 years, after the last product has been placed on the market.
Affix CE marking
Once the necessary steps have been successfully completed, the CE marking must be affixed to the medical device. The CE marking has to be placed visibly and legibly on the product or, if not possible due to the nature of the product, be affixed to the packaging and the accompanying document. The CE marking shall consist of the initials 'CE' taking the following form:
The various components of the CE marking must have the same vertical dimension and may not be smaller than 5 mm. If the CE marking is reduced or enlarged, the proportions given in the graduated drawing above must be respected.
When the product is subject to other Directives covering other aspects and which also provide for the ‘CE’ marking, the accompanying documents must indicate that the product also conforms to those other Directives.
If a Notified Body has been involved in the conformity assessment procedure, its identification number must also be displayed.
△ The above picture and text are from the Internet. If there is any infringement, please contact us in time and we will delete it immediately.
As the first global market access digital platform, BWTCmall.com offers a new and unprecedented online service experience for its many users.
As a technology-driven company, BWTCmall.com is committed to developing a powerful and scalable platform that not only enables global manufacturers to quickly and accurately find the best way to reach the global market, but also to experience cost-effective cutting-edge technology and services. Let the global market access market truly enter a virtuous circle of fairness, objectivity, harmony and win-win.
Please visit the solution for get your best way of testing and certification / approvals to the target market.